Educational simulator — not for clinical use. The tracings are synthetic and, for pathologies, approximate. Do not use for diagnosis, treatment, or any patient-care decision.
What is exact: lead formation (Einthoven/Goldberger), electrode reversals, artifact-to-lead distribution via the Wilson Central Terminal, and filter behaviour all follow the physics directly — that is the whole point of this module.
The underlying ECG is a synthetic normal-sinus waveform (a single cardiac dipole projected through the Dower transform) — realistic but idealised. This is a teaching module about electrodes, cable reversals, artifacts and filters, not a pathology library.
Citations
Mondal S, Raja DP, Muslim NA, Prabhu MA. Electrocardiographic artifacts in clinical practice: a logical approach to recognition and prevention. World J Cardiol 2026;18(3):116299.
Dower GE, et al. Deriving the 12-lead electrocardiogram from four (EASI) electrodes / the Dower transformation. J Electrocardiol 1988.
Kors JA, et al. Reconstruction of the Frank vectorcardiogram from standard leads. Eur Heart J 1990.